AERONET Instrument Troubleshooting Interface - Solutions for LOW AVERAGE CIMEL INTERNAL BATTERY Flag
Description Models 4 and 5 contain an internal 6 V battery that serves as a backup and helps regulate the power supply to internal electronics. The Model T does not have an internal battery and is not subject to this flag.
Condition(s) Model 4.8x and Model 5.x Only: Average Cimel battery - flag if < 5 volts, do not include voltages < 4 volts and > 6 volts.
How To Check/Test Model 4 and 5: VIEW > BAT (disconnect solar panel before checking)
Solution Level 1 (< 5 minutes) Model 4 and 5: 1. See External battery connections Model 5.jpg 2. Check/reseat all connections between solar panel and control box. 3. Check/reseat connector panel.
Solution Level 2 (more involved) Model 4 and 5: 1. Check/reseat internal connections and check TIP32 transistor for burning. See Internal check control box.jpg and Testing TIP32 Transistor.pdf 2. Check power coming directly from solar panel.
Solution Level 3 (last resort) Model 4 and 5: 1. Replace battery or solar panel. 2. Cut out the TIP32 transistor. See TIP burnt bypass.jpg 3. Return instrument for maintenance.